Lux sensor hardware

Glasses-type trackers (Pupil Core, Pupil Neon, Tobii Pro Glasses 3) see the world through a scene camera with automatic exposure, so their video cannot tell how bright the scene is. An Adafruit TSL2591 light sensor mounted on the tracker, facing forward, measures the illuminance at the eye; the analysis turns it into the average luminance of the view (see Luminance).

The firmware and 3D-printable mount are in the repository's Lux Sensor/ folder.

Two ways to log

Logged to a computer Logged to the SD card
Hardware microcontroller + sensor microcontroller + data logging shield (RTC, SD card) + battery + sensor
Recording the sensor logger (or tools/lux_logger.py) over USB power the board on
Clock the computer's, the same as the eye tracker's the board's real-time clock: drifts seconds per day, no time zones
Cable 3–4 m ribbon cable to the computer 1.5 m, logger carried by the participant

Either way the readings (about 10 per second) are logged with the time of each. The sensor logger writes one lux.csv per recording (unix time (s), lux, with a header). The SD card logger and tools/lux_logger.py write one CSV per hour, <month>_<day>_<hour>.csv, with the Unix time in milliseconds in the first column and lux in the fifth. The analysis reads both.

With the SD card logger, set the real-time clock to local time and correct the remaining offset with the time lag parameter (for example, 3600 s for a clock one hour off).

Bill of materials

Tested configurations (the two differ only in their real-time clock, so each has its own firmware):

Adafruit Feather (Lux Sensor/Adalogger FeatherWing/Light_sd_log_direct)

Arduino Nano (Lux Sensor/Arduino Nano/Light_sd_log_direct_nano)

  • Arduino Nano and the Deek-Robot 8105 Nano data logging shield (DS1307 real-time clock)
  • Adafruit TSL2591 sensor and a four-conductor ribbon cable

Only the microcontroller and sensor are needed when always logging to a computer.

Wiring

TSL2591 Microcontroller
Vin (3–5 V) 3V
GND GND
SCL SCL (I²C clock)
SDA SDA (I²C data)

Firmware

Flash the board with the Arduino IDE. The sketches need the Adafruit_Sensor and Adafruit_TSL2591 libraries, and the RTC library for the board's clock (RTC_PCF8523 from RTClib for the FeatherWing, DS1307RTC for the Nano shield). Set the clock once with the RTC library's example sketch.

Adafruit's guides: Feather M4 setup, Adalogger RTC, TSL2591 wiring and test.

Mount

Lux Sensor/Mount Hardware/Adafruit Pupil lumiance sensor kit.stl holds the sensor on the Pupil Core's world camera and includes cable clips. It prints on a generic filament printer.

Calibration

The analysis converts lux to average luminance with

\[ \bar L = \frac{g \cdot E + o}{\Omega} \]

where \(E\) is the reading in lux, \(g\) and \(o\) an optional recalibration of it (default 1 and 0: the chip's own lux), and \(\Omega\) = 2.2 the ratio of illuminance to average luminance for the sensor in its housing (parameters lux_gain, lux_offset, lux_solid_angle). 1.x used \(g\) = 1.706061 and \(o\) = 0.66935; see below for why 2.0 does not.

Where 2.2 comes from. It is not a solid angle, although 1.x called it one: for a uniform field of luminance \(L\), a sensor with angular response \(R(\alpha)\) reads

\[ E = L \int_0^{2\pi}\!\!\int_0^{\theta} R(\alpha)\,\sin\alpha\;d\alpha\,d\varphi \]

over the field it sees. The TSL2591's response is close to a cosine (datasheet, Fig. 12; over a hemisphere the integral is 3.10, against \(\pi\) for an ideal cosine). In the printed housing the board sits at the bottom of the funnel, so the photodiode is level with the funnel's base, 3.7 mm below its rim (10.6 mm across); the rim limits the field to a half-angle of 55°. The integral over that field is 2.12–2.15 (depending on where the chip sits under the funnel), within 2–4 % of 2.2, i.e. about 0.01 mm of modelled pupil.

Where the 1.x gain and offset came from. The 1.x values \(g\) = 1.706061, \(o\) = 0.66935 came from a comparison in the 2019 master thesis (Pignoni, "Quantitative evaluation tool of cognitive workload", NTNU; section 4.2, Table 2 and Figure 13): a projected screen in a dark room was measured by a Konica Minolta CS-2000 spectroradiometer (a 2° spot at its centre, its brightest part) and by the head-mounted sensor (lux / 2.2, a field of more than 60° including the darker surround). Regressing the spectroradiometer on the sensor gives a slope of 1.709 and an intercept of 0.54 cd/m² (r = 0.9998, on the rounded table). The line therefore maps the sensor's field average onto the luminance of the centre of that particular scene; it corrects a field-of-view mismatch, not the sensor. In other scenes the factor differs (about 1 in a uniform field), and with the scene video it would correct twice: the video already turns the field average into the luminance at the gaze. 2.0 therefore uses the chip's lux as reported (about ±15 % between units, datasheet) (open issue 23).