ΔPD and alignment

pipeline.run. Comparing the measured pupil with the expected one.

Offset

The model predicts the pupil of an average observer, so the measured pupil is shifted onto it. alignment sets how the offset is chosen:

Mode Offset Use
recording median of expected − measured over the whole recording default, as in 1.x; ΔPD is centred on zero
baseline the same median, over the events named in baseline_events (default "Riposo, Rest, Baseline") ΔPD relative to rest phases; falls back to the whole recording with a warning if none are found
fixed pupil_offset the value fitted on the participant's calibration sequence, applied unchanged
none 0 absolute comparison

Only valid samples (outside long gaps) enter the median.

ΔPD

\[ \Delta PD = PD_\text{measured} + \text{offset} - PD_\text{expected} \]

averaged over consecutive windows of cw_window (0.2 s), ignoring missing samples, then smoothed with a first-order Savitzky–Golay filter over cw_smoothing windows on each side (1: three windows). Windows without valid samples stay empty.

Statistics

Value Meaning
ΔPD RMS root mean square of ΔPD about zero: how far the pupil strays from the model overall
ΔPD SD standard deviation of ΔPD: its spread around its own mean
ΔPD in SD units ΔPD / SD, exported next to ΔPD in mm
Light left in ΔPD how much of ΔPD the luminance still explains: the slope of ΔPD on log₁₀ luminance (mm per tenfold luminance) and the R² of ΔPD on log₁₀ luminance and its recent change, without the windows within 0.5 s of a luminance step (there the smoothed measured pupil and the expected one differ even for a perfect model). Shown in the summary; a warning above R² 0.1

The light left in ΔPD is the first check of a result: if the model removed the light, ΔPD does not follow the luminance, and both figures are near 0. A large value means ΔPD changes where the light changes may not be workload. It cannot tell model error from a task that itself follows the light (a bright task screen and a dark rest screen): there, judge it on a recording, or a part, where the light varies and the task does not.

Absolute ΔPD in mm depends on the light sensitivity, the display photometry and the offset. Reporting changes relative to a baseline, or in SD units, compares better across participants; Eckert et al. (2022) reach the same conclusion for PLR-corrected pupil sizes.

Events

For each event, the mean ΔPD inside it (in mm and SD units) is written to the events CSV. Events come from the event_log CSV and, for the Neon and the Glasses 3, from the recording's own events.