Participant calibration

Individual differences in pupil size, light sensitivity and response speed, and the imperfectly known photometry of a headset's display, limit how well the expected pupil matches the measured one. For the Varjo XR-4 each participant watches a full-field calibration sequence, which is used in two automatic fits:

  1. Light sensitivity: how strongly this participant's pupil responds to the display's light, and to its red, green and blue channels.
  2. Latency and dynamics: response latency, dilation and constriction speed.

Both start from the display photometry: the headset's nominal black and white luminance and gamma, which you provide (e.g. from the datasheet) and keep in their own file. The fits are saved in the participant's parameter file and applied to their other recordings.

The sequence

The default sequence is the calibration presenter's full calibration: the 8 grey levels of Eckert et al. (2022), longer the darker; red, green and blue at 64, 128, 191 and 255, each after a black step; two greys repeated; 34 steps, 291 s, in a pseudo-random order (calibration.DEFAULT). The presenter usually scrambles it again for each participant, seeded by their ID: load the run file it saves (Load sequence…), which records the order played, or regenerate it with calibration.scrambled(calibration.DEFAULT, "<participant ID>").

The 20-step staircase used before October 2026 (greys 0 to 255, then red, green and blue at 64 to 255, 6 s each, ascending) is still available (Built-in menu, calibration.STAIRCASE_20) for recordings made with it; its fixed ascending order confounds luminance with time (open issues 36, 41, 43). Other sequences are loaded from a CSV:

time,r,g,b,label
0,0,0,0,Black
6,36,36,36,Gray 36
...

Accepted layouts: a header with a time column (time, t, timestamp, start, onset; seconds, or milliseconds if the header says ms) and r/red, g/green, b/blue columns, optionally duration and label; or four unlabelled columns time, R, G, B. Colours may be 0–255 or 0–1. Each step lasts until the next starts; the last uses its duration, or the median step length.

0. Display photometry (once per headset and settings)

In Display photometry (datasheet) enter the headset's black luminance (Lmin), white luminance (Lmax) and gamma, then File → Save display photometry…. The file is remembered for that device and loaded automatically with its recordings; load another one with File → Load display photometry… if the headset's settings or firmware change.

These values do not have to be exact. The sequence cannot tell a dimmer headset from a less sensitive participant, so any error common to all participants ends up in each participant's light sensitivity (see below). What matters is using the same display photometry for all participants and recordings made with the same headset settings.

1. Locate the sequence

Open the recording and click Find in recording. It finds the step changes in the analysed video colour, tries each as a start, and scores how well the sequence's colours explain the measured ones (with one gain per channel, since recorded levels are below nominal). If the steps are evenly spaced it also tries the recording's actual step length, so a sequence played with 10 s steps is found even though the built-in one uses 6 s. Stretches where eye tracking was lost have no analysed colour and are left out of the score; the status bar says how much of the sequence fell in such gaps.

The overlay turns on and the start is set. Adjust it by dragging its handle under the plots if needed.

2. Fit the light sensitivity

Click 1. Fit light sensitivity (tick Also fit gamma only if the datasheet gamma clearly fails). For every step the fit takes the pupil's level at the end of the step (the mean of its last 30 %) and finds the values that make the model match them:

Parameter Determined by
Light sensitivity where along the grey steps the pupil stops shrinking
Red/green/blue weight the size of the responses to the colour steps, relative to the greys
Gamma (optional) the spacing of the grey steps
Black level (optional) the pupil on the black steps, with the white held at its nominal value

The pupil offset is not fitted: with the sensitivity and the display's black and white free, an offset would hide an error of them. It stays the value set by hand. The black level needs steps long enough to dark-adapt: with short steps the black looks brighter than it is, and the fit reports the contrast it implies against the nominal one.

The result window plots the level of each step in its colour against the model before and after, with the sensitivity's 95 % interval and warnings (e.g. when steps were still dilating at their end, or when the sensitivity is weakly determined). Apply sets the values.

A sensitivity of 2 means this participant's pupil responds to the display as the standard observer would to twice the luminance, or equally that the display is twice as bright as its photometry says. Compare sensitivities only between participants calibrated with the same display photometry.

3. Fit latency and dynamics

Click 2. Fit latency and offset (tick Include dynamics (time constants and transient) to fit the dilation and constriction time constants together with the constriction beyond the steady state after brightening steps and its re-dilation, and turn the dynamics on). The latency is measured from the constriction onsets at the sequence's brightening steps (the result reports it with its spread); the rest is fitted on the sequence window, and the ΔPD RMS before and after is reported. Applying it sets:

  • delay (latency),
  • with dynamics: dynamics = on, attack and release (dilation and constriction time constants), transient and escape,

If a value ends on a search limit, a warning explains it: usually the light sensitivity fit needs another look (sequence start, display photometry).

The pupil scale is not fitted

pupil_correction, the multiplier on the device's pupil scale, stays the value you set (1 by default). One sequence does not determine it: it trades off with the light sensitivity and the black point, and a fitted value far from 1 was not evidence of a wrong scale (open issue 1). Set it by hand when there is a reason, such as a known-size reference or a device whose millimetres are known to be off. See

Calibration fit for how it works.

Glasses with a lux sensor

Pupil Core, Neon and Glasses 3 have no display photometry, so the two fits above are not available. With a lux log (and the presenter played on the computer that recorded the tracker), Load sequence… takes the presenter's run file, which places the sequence, and Fit sensitivity and offset fits the light sensitivity and the pupil offset on it: see Calibration fit. Check first that the sensor sees the screen: a sensor dominated by room light barely follows the steps, and the fit says so.

4. Save

File → Save parameters as… and name the file after the participant. Load it with their other recordings, or pass it to cwtool --params (with the display photometry file as --display). Participant files do not contain the display photometry, so loading one keeps the display photometry currently in use. (Files from before this change still carry Lmin, Lmax and gamma; loading them reproduces their original results.)

Light sensitivity first

The latency fit takes the light response as given. Running it first makes the latency and time constants absorb luminance errors: on the April 2026 Varjo sample, default settings gave a latency of 0 s, which is not physiological.

Adjusting by hand

Every fitted value stays editable. Watching the expected curve against the measured one and the ΔPD RMS in sequence while changing the sensitivity or channel weights is still a useful check of the fit.